lmn:isr:theory
Differenze
Queste sono le differenze tra la revisione selezionata e la versione attuale della pagina.
Entrambe le parti precedenti la revisioneRevisione precedenteProssima revisione | Revisione precedente | ||
lmn:isr:theory [30/12/2013 18:14] – davide.orsi | lmn:isr:theory [31/12/2013 10:13] (versione attuale) – davide.orsi | ||
---|---|---|---|
Linea 1: | Linea 1: | ||
+ | ==== Instrument design ==== | ||
+ | {{lmn: | ||
+ | |||
+ | The interfacial rheometer we developed is an upgraded version of the | ||
+ | instrument described by Fuller, adapted to the study of photosensitive | ||
+ | polymers. A sketch and two photographs of the instrument are shown in | ||
+ | the figure aside. | ||
+ | |||
+ | The instrument is build around the Langmuir trough described [[lmn: | ||
+ | long, define a channel used to constrain the needle position along the | ||
+ | direction parallel to the short edge of the trough. The stainless steel | ||
+ | needle is typically $13mm$ long and has a diameter of $0.35mm$; its | ||
+ | surface is kept clean by immersion in chloroform. The needle is | ||
+ | magnetized to saturation value before each experiment by contact with a | ||
+ | permanent magnet. | ||
+ | |||
+ | Two coils in Helmholtz configuration (diameter $D=32cm$, resistance | ||
+ | $R=7\Omega$) are placed around the Langmuir trough: each of them carries | ||
+ | a static current $I_0 = 1 A$. The static magnetic field generated by | ||
+ | $I_0$ aligns the needle along the axis of the coils, as shown in the figure below. The whole system is oriented parallel to the North-South | ||
+ | direction, so the Earth field does not affect the alignment of the | ||
+ | needle. A sinusoidal current is added to one of the coils, producing the | ||
+ | oscillating field gradient that moves the needle. A typical sinusoidal | ||
+ | wave of frequency $0.25Hz$ and amplitude $I=1mA$ generates a force $F$ | ||
+ | of the order of $50nN$ on the needle, resulting in a displacement of | ||
+ | about $10\mu m$. | ||
+ | |||
+ | This “two-coils” design is a simplification of the design originally | ||
+ | proposed by Fuller and coworkers, which makes use of two coils in | ||
+ | Helmholtz configuration to generate the static magnetic field, and two | ||
+ | smaller coils in anti-Helmholtz configuration as the source of the | ||
+ | oscillating magnetic gradient that moves the needle. In the following, | ||
+ | we show that the use of two coils do not alter significantly the | ||
+ | behavior of the oscillating magnetic field gradient, with respect to the | ||
+ | one obtained with four coils. | ||
+ | |||
+ | We name $x$ the distance from the center of the trough, along the axis | ||
+ | of the two coils. Figure below, left panel reports the magnetic fields $H_1$ | ||
+ | and $H_2$ generated by each coil -placed respectively at $x=-8cm$ and | ||
+ | $x=8cm$- and by the two coils in Helmholtz configuration. Close to | ||
+ | $x=0cm$, the total field $H_{tot}$ is constant over $x$. When the | ||
+ | oscillating current $I$ is added to one coil, near to $x=0cm$ the field | ||
+ | $H_{tot}$ grows linearly with $x$. If the needle, aligned parallel to | ||
+ | the axis of the coils, is described as a magnetic dipole of momentum | ||
+ | $\mu$, the force exerted on the needle is given by | ||
+ | |||
+ | $$\overrightarrow{F}(x) = \overrightarrow{\mu} \times \nabla \overrightarrow{H}(x)$$ | ||
+ | |||
+ | In the following, we consider only the component of the force parallel | ||
+ | to the axis of the two coils. A calculation of the behavior of $F(x)$ | ||
+ | for our instrument is shown in the central panel of the next figure. Close to $x=0$, $F$ | ||
+ | is directly proportional to $x$: therefore, an elastic constant can be | ||
+ | defined as | ||
+ | |||
+ | $$k=\frac{F(x)}{x}.$$ | ||
+ | |||
+ | {{lmn: | ||
+ | |||
+ | Given this linearity, the energy of the dipole-field interaction $U(x)$ | ||
+ | is parabolic in proximity of $z=0$. The needle will change its position | ||
+ | accordingly to the position $x_0$ of the minimum of $U$. Since $x_0$ | ||
+ | depends on the value of the oscillating current $I$, we have that | ||
+ | |||
+ | $$x_0 = A e^{i\omega t}.$$ | ||
+ | |||
+ | One may consider that, while in the four-coils configuration the value | ||
+ | $k$ is constant, this do no happen in a two-coils design. | ||
+ | |||
+ | The calculation reported in the previous figures | ||
+ | shows that the dependence of $k$ over the current $I$ is negligible in the | ||
+ | present experimental setup; the relative variation of $k$ reaches a | ||
+ | maximum value of the order of $10^{-3}$. | ||
+ | |||
+ | The resulting displacement $d(\omega)$ of the needle -of the order of | ||
+ | $50\mu m$- is detected by a CCD camera [1] equipped with a long focal | ||
+ | objective [2]. An hardware DAQ board [3] is used to control the | ||
+ | instrument: using its analog outputs it is possible to drive the trough | ||
+ | barriers and to apply the oscillating current to the Helmholtz coils via | ||
+ | a set of power supplies. The DAQ analog inputs are used to measure the | ||
+ | trough area, the surface pressure and the current on the coils as | ||
+ | functions of time. Both the DAQ hardware board and the CCD camera are | ||
+ | triggered by an external reference wave, oscillating at 14Hz. The | ||
+ | external triggering is needed to properly measure the phase lag | ||
+ | $\delta(\omega)$ between stress and strain, ensuring that the correct | ||
+ | mechanical response is retrieved. A connection scheme of the | ||
+ | experimental setup is sketched in the figure below: the connections relative to the trigger signal are depicted in | ||
+ | red. | ||
+ | |||
+ | - The Imaging Source, 1024x768 pixels | ||
+ | - Mitutoyo MPLAN-APO-20X | ||
+ | - National Instruments PCI-6036e | ||
+ | |||
+ | {{lmn: | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | |||
+ | ==== Calibration: | ||
+ | |||
+ | In order to characterize the inertial response of the needle, we | ||
+ | consider the oscillatory movement of the needle on a pure water surface. | ||
+ | The total force exerted on the needle given by the composition of: | ||
+ | |||
+ | **a)** a viscous drag force $F_c = -d \dot{x}$, where $d$ is the drag coefficient for a motion parallel to the axis of the needle. If the needle is thin - its radius is much shorter than its length ($a \ll L$)- then the drag coefficient is given by a relatively simple expression @Levine2004 | ||
+ | |||
+ | $$d= \frac{2 \pi \eta L}{\log \left( 0.43 L/l_0 \right)}$$ | ||
+ | |||
+ | The drag coefficient $d$ is then related to the water viscosity $\eta$, the needle length $L$ and a characteristic length $l_0$ over which the two dimensional fluid velocity field can vary, which can be safely assumed to be of the order of the needle’s radius $a$; | ||
+ | |||
+ | **b)** an harmonic force originating from the interaction between the magnetic dipole of the needle and the magnetic field gradient generated by the system of coils, | ||
+ | |||
+ | $$F_m = -k \left( x-x_0 \right) = - \frac{\partial^2 U}{\partial x^2} \left( x-x_0 \right)$$ | ||
+ | |||
+ | where k is the elastic constant previously defined and U is the energy of the dipole-field interaction. The center $x_0$ of the harmonic potential oscillates with angular frequency $\omega$ (see equation [eq: | ||
+ | |||
+ | The resulting differential equation is | ||
+ | |||
+ | $$m \ddot{x} = - d\dot{x} -kx + kAe^{i\omega t}$$ | ||
+ | |||
+ | This equation is that of a forced-dumped oscillator of mass $m$, | ||
+ | spring constant $k$ and damping constant $d$ @Reynaert2008. The ratio | ||
+ | $AR$ of the amplitudes of the needle and force oscillation, | ||
+ | phase lag $\delta$ result as: | ||
+ | |||
+ | $$AR=\frac{X}{kA}=\frac{1}{\sqrt{(k-m\omega^2)^2+(\omega d)^2}}$$ | ||
+ | |||
+ | $$\delta=\arctan \left( \frac{-\omega d}{k-m \omega^2} \right)$$ | ||
+ | |||
+ | Commonly, instead of measuring the elastic constant $k$, the simpler | ||
+ | approach of measuring the oscillatory voltage $V$ applied to one of the | ||
+ | two coils is preferred. Therefore, a calibration constant has to | ||
+ | determined; in order to do that, the amplitude ratio $AR_V=\frac{X}{V}$ | ||
+ | is measured. Observing the high-frequency limit of $AR$, it can be noted | ||
+ | that it is proportional to $\omega^2$ via the mass of the needle $m$. | ||
+ | |||
+ | $$\frac{X(\omega)}{F(\omega)} = \frac{X(\omega)}{\alpha V(\omega)} \rightarrow {m \omega^2}, \; \omega \rightarrow \infty$$ | ||
+ | |||
+ | This relation -in analogy with @Brooks1999- allow the determination of | ||
+ | the *calibration constant* $\alpha$: | ||
+ | |||
+ | $$\alpha = \frac{1}{m\omega^2}\lim_{\omega \rightarrow \infty}{\frac{X}{V}}$$ | ||
+ | |||
+ | A measurement of $AR$ and $\delta$ is reported in the figure below in logarithmic and semilogarithmic scale, respectively. | ||
+ | |||
+ | {{lmn: | ||
+ | {{lmn: | ||
+ | |||
+ | While the mass $m$ of the needle is measured, the oscillator parameters | ||
+ | $k$ and $d$ were determined by a fit performed on the stress-stain ratio | ||
+ | curve using equation [eq: | ||
+ | [tab:kmd]. The phase lag $\delta$ is in very good agreement with the | ||
+ | model curve built from the parameters of table [tab:kmd] (figure | ||
+ | [fig: | ||
+ | |||
+ | ^ $k$ | %% $1.13 \cdot 10^{-5} N/m$ %% | | ||
+ | ^ $m$ | %% $ 9.9 \cdot 10^{-6} Kg$ %% | | ||
+ | ^ $d$ | %% $1.74 \cdot 10^{-5} Kg/s$ %% | | ||
+ | |||
+ | It is worth comparing the value of $d$ obtained from the fit of the | ||
+ | experimental curves with its theoretical value calculated using equation | ||
+ | [eq:drag]. If the characteristic length $l_0$ is taken equal to the | ||
+ | needle’s radius $a$, we obtain $d_t=2.38 \cdot 10^{-5} Kg/s$, with | ||
+ | respect to the measured value of $1.74 \cdot 10^{-5} Kg/s$: the order of | ||
+ | magnitude of the drag is correctly predicted. | ||
+ | |||
+ | |||
+ | ==== Oscillations in presence of a film ==== | ||
+ | |||
+ | |||
+ | |||
+ | The presence of a film at the air/water interface induces an additional | ||
+ | hindrance of the motion of the needle, thus reducing the amplitude of | ||
+ | its oscillations and modifying the phase lag between stress and strain. | ||
+ | |||
+ | An measurement of the response of a Langmuir film can be considered | ||
+ | meaningful is it is significantly different from the measurement | ||
+ | performed in absence of film, which characterize the intrinsic | ||
+ | oscillating behavior of the system. If this deviation is too small, than | ||
+ | the measurement is likely to include huge effects due to the drag of the | ||
+ | subphase. | ||
+ | |||
+ | As stated at the beginning of this chapter, the contribution to the drag | ||
+ | due to the water subphase are negligible if the Boussinesq number $B$ is | ||
+ | high: $B>> | ||
+ | |||
+ | Usually, a so-called *experimental Boussinesq number* @Brooks1999 | ||
+ | [@Reynaert2008] | ||
+ | |||
+ | $$B_{exp} = \frac{AR_{sys}}{AR_{film}}$$ | ||
+ | |||
+ | is used in the literature to quantify the deviation of the response of | ||
+ | the needle from the behavior observed on pure water. If | ||
+ | $B_{exp} > 10-100$ in the frequency range investigated, | ||
+ | mechanical response measured may be considered as not affected by | ||
+ | hydrodynamic contributions due to the bulk of the subphase. In this | ||
+ | case, the shear modulus $G$ is given by | ||
+ | |||
+ | $$G(\omega) = \frac{W}{2L}\frac{\alpha V(\omega)}{X(\omega)} e^{i\delta}$$ | ||
+ | |||
+ | If $B \simeq 1$, the subphase drag contribution to $G$ has to be | ||
+ | decoupled from the one due to the film. A simple way to decouple the | ||
+ | contributions to $G$ arising from the drag of the film and from the drag | ||
+ | of the subphase, when the latter is not negligible, is to imply a linear | ||
+ | relation between the two quantities @Brooks1999a [@Reynaert2008], | ||
+ | that | ||
+ | |||
+ | $$G_{film}= G_{meas}- G^*_{water} =G_{meas} - \frac{W}{2L}AR_{w} \exp{(i\delta_{w})}.$$ | ||
+ | |||
+ | The accuracy of this procedure is questionable. In this work, the | ||
+ | preferred approach consists in setting a confidence threshold value | ||
+ | $B_{exp} = 10$, above which the measurement can be considered as not | ||
+ | influenced by subphase drag contributions. | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | |||
+ | ==== Measuring the response of a Newtonian fluid ==== | ||
+ | |||
+ | An additional control of the instrument calibration is conducted | ||
+ | measuring the response of films of poly(dimethylsiloxane) of controlled | ||
+ | thickness $d$ and known viscosity $\eta_ {bulk}=0.97 Pa\cdot s$. This | ||
+ | oil has the mechanical response of an ideal (newtonian) fluid: it | ||
+ | presents a purely viscous response with frequency-independent shear | ||
+ | viscosity $\eta_s$. | ||
+ | |||
+ | In case of predominantly viscous systems, the generalized viscosity | ||
+ | $\eta_{2D}$ is obtained from $G$ through the simple relation: | ||
+ | |||
+ | $$\eta_{2D}=\frac{G" | ||
+ | |||
+ | It is reasonable to assume that $\eta_{2D}$ is proportional to the | ||
+ | bulk viscosity through the thickness of the film, | ||
+ | $\eta_{2D}= d \cdot \eta_{bulk}$, | ||
+ | |||
+ | {{lmn: | ||
+ | |||
+ | |||
+ | Data are shown in figure above, left panel: the film behaves as a Newtonian | ||
+ | fluid with $G"$ (circles) growing linearly with the frequency, as | ||
+ | indicated by fit represented by the dashed line, while $G'$ is zero | ||
+ | within the error. On the right we compare the value of the | ||
+ | viscosity $\eta_{2D}$ obtained from $G'' | ||
+ | its expected value based on the bulk viscosity. | ||
+ | |||
+ | The results are consistent with the theoretical behavior for an ideal | ||
+ | viscous fluid. However, careful inspection of the data from the lowest | ||
+ | thickness ($d< | ||
+ | shows a small deviation which can be ascribed to the drag due to the | ||
+ | water subphase, which becomes comparable to that of the film when this | ||
+ | becomes too thin. This effect has already been found and discussed in | ||
+ | the literature @Reynaert2008. If we suppose, in a simple approximation, | ||
+ | a linear superposition of these effects, a rough estimate of the | ||
+ | subphase drag is obtained: $|G|_{sub} \simeq 10 \mu N/m$. | ||
+ | |||
+ | ==== Adaptation for photosensitive polymers ==== | ||
+ | |||
+ | |||
+ | The study of the mechanical properties of photosensitive polymers | ||
+ | requires a proper adaptation of the instrument described so far. In | ||
+ | particular, the instrument had to be enclosed in an aluminum box | ||
+ | equipped with removable walls, to ensure that spurious light from the | ||
+ | ambient does not influence the rheological properties of the film. | ||
+ | Moreover, the illumination light required for the tracking of the | ||
+ | needle’s position has to be chosen in a spectral region where the sample | ||
+ | is characterized by low absorbance. For the photosensitive azopolymers | ||
+ | investigated in this work, red light ($\lambda >630nm$) was chosen. | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | |||
+ | < | ||
+ | <script type=" | ||
+ | MathJax.Hub.Config({ | ||
+ | tex2jax: {inlineMath: | ||
+ | }); | ||
+ | </ | ||
+ | <script type=" | ||
+ | </ |