Strumenti Utente

Strumenti Sito


lmn:isr:theory

Differenze

Queste sono le differenze tra la revisione selezionata e la versione attuale della pagina.

Link a questa pagina di confronto

Entrambe le parti precedenti la revisioneRevisione precedente
Prossima revisione
Revisione precedente
lmn:isr:theory [30/12/2013 18:52] davide.orsilmn:isr:theory [31/12/2013 10:13] (versione attuale) davide.orsi
Linea 1: Linea 1:
 +==== Instrument design ====
  
 +{{lmn:isr:isr.png |ISR setup}}
 +
 +The interfacial rheometer we developed is an upgraded version of the
 +instrument described by Fuller, adapted to the study of photosensitive
 +polymers. A sketch and two photographs of the instrument are shown in
 +the figure aside.
 +
 +The instrument is build around the Langmuir trough described [[lmn:langmuir|here]]. Two parallel glass plates, each of them $10cm$
 +long, define a channel used to constrain the needle position along the
 +direction parallel to the short edge of the trough. The stainless steel
 +needle is typically $13mm$ long and has a diameter of $0.35mm$; its
 +surface is kept clean by immersion in chloroform. The needle is
 +magnetized to saturation value before each experiment by contact with a
 +permanent magnet.
 +
 +Two coils in Helmholtz configuration (diameter $D=32cm$, resistance
 +$R=7\Omega$) are placed around the Langmuir trough: each of them carries
 +a static current $I_0 = 1 A$. The static magnetic field generated by
 +$I_0$ aligns the needle along the axis of the coils, as shown in the figure below. The whole system is oriented parallel to the North-South
 +direction, so the Earth field does not affect the alignment of the
 +needle. A sinusoidal current is added to one of the coils, producing the
 +oscillating field gradient that moves the needle. A typical sinusoidal
 +wave of frequency $0.25Hz$ and amplitude $I=1mA$ generates a force $F$
 +of the order of $50nN$ on the needle, resulting in a displacement of
 +about $10\mu m$.
 +
 +This “two-coils” design is a simplification of the design originally
 +proposed by Fuller and coworkers, which makes use of two coils in
 +Helmholtz configuration to generate the static magnetic field, and two
 +smaller coils in anti-Helmholtz configuration as the source of the
 +oscillating magnetic gradient that moves the needle. In the following,
 +we show that the use of two coils do not alter significantly the
 +behavior of the oscillating magnetic field gradient, with respect to the
 +one obtained with four coils.
 +
 +We name $x$ the distance from the center of the trough, along the axis
 +of the two coils. Figure below, left panel reports the magnetic fields $H_1$
 +and $H_2$ generated by each coil -placed respectively at $x=-8cm$ and
 +$x=8cm$- and by the two coils in Helmholtz configuration. Close to
 +$x=0cm$, the total field $H_{tot}$ is constant over $x$. When the
 +oscillating current $I$ is added to one coil, near to $x=0cm$ the field
 +$H_{tot}$ grows linearly with $x$. If the needle, aligned parallel to
 +the axis of the coils, is described as a magnetic dipole of momentum
 +$\mu$, the force exerted on the needle is given by
 +
 +$$\overrightarrow{F}(x) = \overrightarrow{\mu} \times \nabla \overrightarrow{H}(x)$$
 +
 +In the following, we consider only the component of the force parallel
 +to the axis of the two coils. A calculation of the behavior of $F(x)$
 +for our instrument is shown in the central panel of the next figure. Close to $x=0$, $F$
 +is directly proportional to $x$: therefore, an elastic constant can be
 +defined as
 +
 +$$k=\frac{F(x)}{x}.$$
 +
 +{{lmn:isr:isrcampi.png|ISR - magnetic fields}} {{lmn:isr:isrforza.png|ISR - forces}} {{lmn:isr:gradient_variation.png|ISR - gradient variation}}
 +
 +Given this linearity, the energy of the dipole-field interaction $U(x)$
 +is parabolic in proximity of $z=0$. The needle will change its position
 +accordingly to the position $x_0$ of the minimum of $U$. Since $x_0$
 +depends on the value of the oscillating current $I$, we have that
 +
 +$$x_0 = A e^{i\omega t}.$$
 +
 +One may consider that, while in the four-coils configuration the value
 +$k$ is constant, this do no happen in a two-coils design.
 +
 +The calculation reported in the previous figures 
 +shows that the dependence of $k$ over the current $I$ is negligible in the
 +present experimental setup; the relative variation of $k$ reaches a
 +maximum value of the order of $10^{-3}$.
 +
 +The resulting displacement $d(\omega)$ of the needle -of the order of
 +$50\mu m$- is detected by a CCD camera [1] equipped with a long focal
 +objective [2]. An hardware DAQ board [3] is used to control the
 +instrument: using its analog outputs it is possible to drive the trough
 +barriers and to apply the oscillating current to the Helmholtz coils via
 +a set of power supplies. The DAQ analog inputs are used to measure the
 +trough area, the surface pressure and the current on the coils as
 +functions of time. Both the DAQ hardware board and the CCD camera are
 +triggered by an external reference wave, oscillating at 14Hz. The
 +external triggering is needed to properly measure the phase lag
 +$\delta(\omega)$ between stress and strain, ensuring that the correct
 +mechanical response is retrieved. A connection scheme of the
 +experimental setup is sketched in the figure below: the connections relative to the trigger signal are depicted in
 +red.
 +
 +  - The Imaging Source, 1024x768 pixels
 +  - Mitutoyo MPLAN-APO-20X
 +  - National Instruments PCI-6036e
 +
 +{{lmn:isr:schem-isr.png?600px|Electrical connections}}
 +
 +
 +
 +
 +
 +==== Calibration: oscillations on pure water ====
 +
 +In order to characterize the inertial response of the needle, we
 +consider the oscillatory movement of the needle on a pure water surface.
 +The total force exerted on the needle given by the composition of:
 +
 +**a)** a viscous drag force $F_c = -d \dot{x}$, where $d$ is the drag coefficient for a motion parallel to the axis of the needle. If the needle is thin - its radius is much shorter than its length ($a \ll L$)- then the drag coefficient is given by a relatively simple expression @Levine2004
 +
 +$$d= \frac{2 \pi \eta L}{\log \left( 0.43 L/l_0 \right)}$$
 +
 +The drag coefficient $d$ is then related to the water viscosity $\eta$, the needle length $L$ and a characteristic length $l_0$ over which the two dimensional fluid velocity field can vary, which can be safely assumed to be of the order of the needle’s radius $a$;
 +
 +**b)** an harmonic force originating from the interaction between the magnetic dipole of the needle and the magnetic field gradient generated by the system of coils,
 +
 +$$F_m = -k \left( x-x_0 \right) = - \frac{\partial^2 U}{\partial x^2} \left( x-x_0 \right)$$
 +
 +where k is the elastic constant previously defined and U is the energy of the dipole-field interaction. The center $x_0$ of the harmonic potential oscillates with angular frequency $\omega$ (see equation [eq:cenharm]).
 +
 +The resulting differential equation is
 +
 +$$m \ddot{x} = - d\dot{x} -kx + kAe^{i\omega t}$$
 +
 +This equation is that of a forced-dumped oscillator of mass $m$,
 +spring constant $k$ and damping constant $d$ @Reynaert2008. The ratio
 +$AR$ of the amplitudes of the needle and force oscillation, and the
 +phase lag $\delta$ result as:
 +
 +$$AR=\frac{X}{kA}=\frac{1}{\sqrt{(k-m\omega^2)^2+(\omega d)^2}}$$
 +
 +$$\delta=\arctan \left( \frac{-\omega d}{k-m \omega^2} \right)$$
 +
 +Commonly, instead of measuring the elastic constant $k$, the simpler
 +approach of measuring the oscillatory voltage $V$ applied to one of the
 +two coils is preferred. Therefore, a calibration constant has to
 +determined; in order to do that, the amplitude ratio $AR_V=\frac{X}{V}$
 +is measured. Observing the high-frequency limit of $AR$, it can be noted
 +that it is proportional to $\omega^2$ via the mass of the needle $m$.
 +
 +$$\frac{X(\omega)}{F(\omega)} = \frac{X(\omega)}{\alpha V(\omega)} \rightarrow {m \omega^2}, \; \omega \rightarrow \infty$$
 +
 +This relation -in analogy with @Brooks1999- allow the determination of
 +the *calibration constant* $\alpha$:
 +
 +$$\alpha = \frac{1}{m\omega^2}\lim_{\omega \rightarrow \infty}{\frac{X}{V}}$$
 +
 +A measurement of $AR$ and $\delta$ is reported in the figure below in logarithmic and semilogarithmic scale, respectively.
 +
 +{{lmn:isr:calibrazione_acqua_ar.png|Measurement on water - Amplitude ratio}}
 +{{lmn:isr:calibrazione_acqua_del.png|Measurement on water - Phase shift}}
 +
 +While the mass $m$ of the needle is measured, the oscillator parameters
 +$k$ and $d$ were determined by a fit performed on the stress-stain ratio
 +curve using equation [eq:isr-amplratio]; they are reported in table
 +[tab:kmd]. The phase lag $\delta$ is in very good agreement with the
 +model curve built from the parameters of table [tab:kmd] (figure
 +[fig:isr-water]b) using equation [eq:isr-phase].
 +
 +^ $k$ | %% $1.13 \cdot 10^{-5} N/m$  %% |
 +^ $m$ | %% $ 9.9 \cdot 10^{-6} Kg$   %% |
 +^ $d$ | %% $1.74 \cdot 10^{-5} Kg/s$ %% |
 +
 +It is worth comparing the value of $d$ obtained from the fit of the
 +experimental curves with its theoretical value calculated using equation
 +[eq:drag]. If the characteristic length $l_0$ is taken equal to the
 +needle’s radius $a$, we obtain $d_t=2.38 \cdot 10^{-5} Kg/s$, with
 +respect to the measured value of $1.74 \cdot 10^{-5} Kg/s$: the order of
 +magnitude of the drag is correctly predicted.
 +
 +
 +==== Oscillations in presence of a film ====
 +
 +
 +
 +The presence of a film at the air/water interface induces an additional
 +hindrance of the motion of the needle, thus reducing the amplitude of
 +its oscillations and modifying the phase lag between stress and strain.
 +
 +An measurement of the response of a Langmuir film can be considered
 +meaningful is it is significantly different from the measurement
 +performed in absence of film, which characterize the intrinsic
 +oscillating behavior of the system. If this deviation is too small, than
 +the measurement is likely to include huge effects due to the drag of the
 +subphase.
 +
 +As stated at the beginning of this chapter, the contribution to the drag
 +due to the water subphase are negligible if the Boussinesq number $B$ is
 +high: $B>>1$.
 +
 +Usually, a so-called *experimental Boussinesq number* @Brooks1999
 +[@Reynaert2008]
 +
 +$$B_{exp} = \frac{AR_{sys}}{AR_{film}}$$
 +
 +is used in the literature to quantify the deviation of the response of
 +the needle from the behavior observed on pure water. If
 +$B_{exp} > 10-100$ in the frequency range investigated, then the
 +mechanical response measured may be considered as not affected by
 +hydrodynamic contributions due to the bulk of the subphase. In this
 +case, the shear modulus $G$ is given by
 +
 +$$G(\omega) = \frac{W}{2L}\frac{\alpha V(\omega)}{X(\omega)} e^{i\delta}$$
 +
 +If $B \simeq 1$, the subphase drag contribution to $G$ has to be
 +decoupled from the one due to the film. A simple way to decouple the
 +contributions to $G$ arising from the drag of the film and from the drag
 +of the subphase, when the latter is not negligible, is to imply a linear
 +relation between the two quantities @Brooks1999a [@Reynaert2008], so
 +that
 +
 +$$G_{film}= G_{meas}- G^*_{water} =G_{meas} - \frac{W}{2L}AR_{w} \exp{(i\delta_{w})}.$$
 +
 +The accuracy of this procedure is questionable. In this work, the
 +preferred approach consists in setting a confidence threshold value
 +$B_{exp} = 10$, above which the measurement can be considered as not
 +influenced by subphase drag contributions.
 +
 +
 +
 +
 +
 +==== Measuring the response of a Newtonian fluid ====
 +
 +An additional control of the instrument calibration is conducted
 +measuring the response of films of poly(dimethylsiloxane) of controlled
 +thickness $d$ and known viscosity $\eta_ {bulk}=0.97 Pa\cdot s$. This
 +oil has the mechanical response of an ideal (newtonian) fluid: it
 +presents a purely viscous response with frequency-independent shear
 +viscosity $\eta_s$.
 +
 +In case of predominantly viscous systems, the generalized viscosity
 +$\eta_{2D}$ is obtained from $G$ through the simple relation:
 +
 +$$\eta_{2D}=\frac{G"}{\omega}.$$
 +
 +It is reasonable to assume that $\eta_{2D}$ is proportional to the
 +bulk viscosity through the thickness of the film,
 +$\eta_{2D}= d \cdot \eta_{bulk}$, thus ignoring confinement effects.
 +
 +{{lmn:isr:calibrazione1.png| viscous response of a Newtonian fluid}}{{lmn:isr:calibrazione2.png| Sensitivity}}
 +
 +
 +Data are shown in figure above, left panel: the film behaves as a Newtonian
 +fluid with $G"$ (circles) growing linearly with the frequency, as
 +indicated by fit represented by the dashed line, while $G'$ is zero
 +within the error. On the right we compare the value of the
 +viscosity $\eta_{2D}$ obtained from $G''$ for each film thickness d with
 +its expected value based on the bulk viscosity.
 +
 +The results are consistent with the theoretical behavior for an ideal
 +viscous fluid. However, careful inspection of the data from the lowest
 +thickness ($d<50\mu m$, represented by the gray area in the picture)
 +shows a small deviation which can be ascribed to the drag due to the
 +water subphase, which becomes comparable to that of the film when this
 +becomes too thin. This effect has already been found and discussed in
 +the literature @Reynaert2008. If we suppose, in a simple approximation,
 +a linear superposition of these effects, a rough estimate of the
 +subphase drag is obtained: $|G|_{sub} \simeq 10 \mu N/m$.
 +
 +==== Adaptation for photosensitive polymers ====
 +
 +
 +The study of the mechanical properties of photosensitive polymers
 +requires a proper adaptation of the instrument described so far. In
 +particular, the instrument had to be enclosed in an aluminum box
 +equipped with removable walls, to ensure that spurious light from the
 +ambient does not influence the rheological properties of the film.
 +Moreover, the illumination light required for the tracking of the
 +needle’s position has to be chosen in a spectral region where the sample
 +is characterized by low absorbance. For the photosensitive azopolymers
 +investigated in this work, red light ($\lambda >630nm$) was chosen.
 +
 +
 +
 +
 +
 +<html>
 +<script type="text/x-mathjax-config">
 +MathJax.Hub.Config({
 +  tex2jax: {inlineMath: [['$','$'], ['\\(','\\)']]}
 +});
 +</script>
 +<script type="text/javascript" src="http://cdn.mathjax.org/mathjax/latest/MathJax.js?config=TeX-AMS-MML_HTMLorMML"></script>
 +</html>

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki